Deprecated: Creation of dynamic property Seboettg\CiteProc\Root\Info::$title-short is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/citationStyleLanguage/lib/vendor/seboettg/citeproc-php/src/Root/Info.php on line 62
Deprecated: Creation of dynamic property Seboettg\CiteProc\Root\Info::$category is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/citationStyleLanguage/lib/vendor/seboettg/citeproc-php/src/Root/Info.php on line 62
Deprecated: Creation of dynamic property Seboettg\CiteProc\Root\Info::$updated is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/citationStyleLanguage/lib/vendor/seboettg/citeproc-php/src/Root/Info.php on line 62
Deprecated: Creation of dynamic property Seboettg\CiteProc\Root\Info::$rights is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/citationStyleLanguage/lib/vendor/seboettg/citeproc-php/src/Root/Info.php on line 62
Deprecated: htmlspecialchars(): Passing null to parameter #1 ($string) of type string is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/googleScholar/GoogleScholarPlugin.php on line 120
Deprecated: htmlspecialchars(): Passing null to parameter #1 ($string) of type string is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/googleScholar/GoogleScholarPlugin.php on line 120
Deprecated: htmlspecialchars(): Passing null to parameter #1 ($string) of type string is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/googleScholar/GoogleScholarPlugin.php on line 120
Deprecated: htmlspecialchars(): Passing null to parameter #1 ($string) of type string is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/googleScholar/GoogleScholarPlugin.php on line 120
Deprecated: htmlspecialchars(): Passing null to parameter #1 ($string) of type string is deprecated in /var/www/vhosts/revistadefisicamedica.es/httpdocs/plugins/generic/googleScholar/GoogleScholarPlugin.php on line 120
Comparative study of patient doses calculated with two methods for breast digital tomosynthesis
| Revista de Física Médica
Comparative study of patient doses calculated with two methods for breast digital tomosynthesis
Authors
María Castillo
Margarita Chevalier
Alfonso Calzado
Julia Garayoa
Julio Valverde
Abstract
In this study, the average glandular doses ( D<sub>G</sub> ) delivered in breast tomosynthesis examinations were estimated over a sample of 150 patients using two different methods. In method 1, the conversion factors air-kerma to D<sub>G</sub> used were those tabulated by Dance et al. and in method 2 were the ones from Feng et al. The protocol for the examination followed in the unit of this study consists in two views per breast, each view composed by a 2D acquisition and a tomosynthesis scan (3D). The resulting D<sub>G</sub> values from both methods present statistically significant differences ( p = 0.02 ) for the 2D modality and were similar for the 3D scan ( p = 0.22 ). The estimated median value of D<sub>G</sub> for the most frequent breasts (thicknesses between 50 and 60 mm) delivered in a single 3D acquisition is 1.7 mGy (36% and 17% higher than the value for the 2D mode estimated with each method) which lies far below the tolerances established by the Spanish Protocol Quality Control in Radiodiagnostic (2011). The total D<sub>G</sub> for a tomosynthesis examination (6.0 mGy) is a factor 2.4 higher than the dose delivered in a 2D examination with two views (method 1).