El mamógrafo Fuji Amulet 1000 MS y la tasa de dosis de radiación en el receptor de imagen
| Revista de Física Médica
Durante las medidas realizadas en las pruebas de aceptación y de estado de referencia del mamógrafo digital Fuji Amulet 1000 MS se obtuvieron resultados contradictorios con respecto al cumplimiento de las tolerancias definidas para la tasa de dosis a la entrada del receptor en el Protocolo Español de Control de Calidad en Radiodiagnóstico. Se comprobó que este resultado se debe al carácter pulsado de la radiación emitida por el tubo y se concluye recomendar que en estos casos se mida la tasa de dosis realizando disparos con valores de carga típicos de exploraciones clínicas (50 mAs o superiores).