Comparison of step and shoot IMRT treatment plans generated by three inverse treatment planning systems

Authors

  • JM Pérez Moreno
  • D Zucca Aparicio
  • P Fernández Letón
  • J García Ruiz-Zorrilla
  • A Miñambres Moro

Abstract

One of the most important issues of intensity modulated radiation therapy (IMRT) treatments using the “step-and-shoot” technique is the number of segments and monitor units (MU) for treatment delivery. These parameters depend heavily on the inverse optimization module of the treatment planning system (TPS) used.

Three commercial treatment planning systems: CMS XiO, iPlan and Prowess Panther have been evaluated. With each of them we have generated a treatment plan for the same group of patients, corresponding to clinical cases. Dosimetric results, MU calculated and number of segments were compared.

Prowess treatment planning system generates plans with a number of segments significantly lower than other systems, while MU are less than a half. It implies important reductions in leakage radiation and delivery time. Degradation in the final dose calculation of dose is very small, because it directly optimizes positions of multileaf collimator (MLC).

Downloads

Issue

Section

Technical note

How to Cite

Comparison of step and shoot IMRT treatment plans generated by three inverse treatment planning systems. (2011). Revista De Física Médica, 12(2). https://revistadefisicamedica.es/index.php/rfm/article/view/56
Bookmark and Share