Two commercial intraoperative gamma probes response to different measurement conditions: comparative analysis

Authors

  • C Camacho López Sección de Radiofísica. Servicio de Medicina Nuclear. Hospital Universitario La Fe. Valencia.
  • JF Martí Vidal Sección de Radiofísica. Servicio de Medicina Nuclear. Hospital Universitario La Fe. Valencia.
  • M Falgás Lacueva Servicio de Medicina Nuclear. Hospital Universitario La Fe. Valencia.
  • C Olivas Arroyo Unidad de Radiofarmacia. Servicio de Medicina Nuclear. Hospital Universitario La Fe. Valencia.

Abstract

Intraoperative gamma radiation probes are increasingly used for localization of sentinel lymph node. Some technical parameters, such as directionality, sensitivity and dead-time effects, which determine the performance to various measurement conditions of two probes, are analyzed in this paper. The sensitivity strongly depends on the source-probe distance, and the type of probe used. It is also shown a suitable angular and spatial resolution, and not noticeable dead time effects are expected on clinical activity levels. The presented tests are simple and reproducible in establishing regular quality controls.