Study of ocular shielding in high dose rate brachytherapy for nasal skin lesions using Monte Carlo simulation and experimental measurements

Authors

  • Gerard Pons Polo Servicio de Protección Radiológica y Radiofísica. Hospital Universitario de Badajoz. España.
  • Pedro Almendral Manzano Servicio de Protección Radiológica y Radiofísica. Hospital Universitario de Badajoz. España.

DOI:

https://doi.org/10.37004/sefm/2026.27.2.A003

Keywords:

Brachytherapy, ocular shielding, lead attenuation, Monte Carlo simulation, EBT 4 films

Abstract

This study evaluates the effectiveness of 2‑mmPb and 4‑mmPb shields for eye protection during high dose rate brachytherapy procedures for nasal skin lesions using an iridium‑192 Flexisource. A realistic clinical scenario was reproduced in an anthropomorphic skull phantom, combining Monte Carlo simulations with the PenRed code and experimental measurements using Gafchromic EBT‑4 films. Transmission factors obtained for the 2‑mmPb and 4‑mmPb shields in the lens region were 0.4 and 0.3 in the simulations, and 0.5 and 0.3 in the experimental measurements, respectively. The simulated total doses to the lens were 9.7 Gy without shielding, 3.9 Gy with 2‑mmPb, and 2.4 Gy with 4‑mmPb. Characterization of secondary electrons emitted by lead showed an 80 % increase in surface dose with a maximum range of 0.8 mm. These results confirm that a 4‑mmPb thickness provides adequate shielding, provided it is accompanied by a coating to prevent surface overdosing.

References

1. Guinot JL, Rembielak A, Perez-Calatayud J, Rodríguez-Villalba S, Skowronek J, Tagliaferri L, et al. GEC-ESTRO ACROP recommendations in skin brachytherapy. Radiother Oncol. 2018;126(3):377-85. https://doi.org/10.1016/j.radonc.2018.01.013 DOI: https://doi.org/10.1016/j.radonc.2018.01.013

2. Jacob S, Michel M, Brézin AP, Laurier D, Bernier MO. Ionizing Radiation as a Risk Factor for Cataract: What about Low-Dose Effects? J Clin Exp Ophthalmol. 2012;S1:005. https://doi.org/10.4172/2155-9570.S1-005 DOI: https://doi.org/10.4172/2155-9570.S1-005

3. Stewart FA, Akleyev AV, Hauer-Jensen M, Hendry JH, Kleiman NJ, MacVittie TJ, et al. ICRP publication 118: ICRP Statement on Tissue Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs — Threshold Doses for Tissue Reactions in a Radiation Protection Context. Ann ICRP. 2012;41(1-2):1-322. https://doi.org/10.1016/j.icrp.2012.02.001 DOI: https://doi.org/10.1016/j.icrp.2012.02.001

4. Zwierzchowski G, Bieleda G, Szymbor A, Boehlke M. Personalized Superficial HDR Brachytherapy—Dosimetric Verification of Dose Distribution with Lead Shielding of Critical Organs in the Head and Neck Region. J Pers Med. 2022;12(9):1432. https://doi.org/10.3390/jpm12091432 DOI: https://doi.org/10.3390/jpm12091432

5. Giménez-Alventosa V, Giménez Gómez V, Oliver S. PenRed: An extensible and parallel Monte-Carlo framework for radiation transport based on PENELOPE. Comput Phys Commun. 2021;267:108065. https://doi.org/10.1016/j.cpc.2021.108065 DOI: https://doi.org/10.1016/j.cpc.2021.108065

6. Computerized Imaging Reference Systems, Inc. Phantom Patient for Stereotactic End-to-End Verification. Norfolk, VA; 2013. Publication No. 038 PB 032423, 16 pp.

7. Granero D, Pérez-Calatayud J, Casal E, Ballester F, Venselaar J. A dosimetric study on the high dose rate Flexisource. Med Phys. 2006;33(12):4578-82. https://doi.org/10.1118/1.2388154 DOI: https://doi.org/10.1118/1.2388154

8. Rivard MJ, Coursey BM, DeWerd LA, Hanson WF, Huq MS, Ibbott GS, et al. Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31(3):633-74. https://doi.org/10.1118/1.1646040 DOI: https://doi.org/10.1118/1.1646040

9. Perez-Calatayud J, Ballester F, Das RK, DeWerd LA, Ibbott GS, Meigooni AS, et al. Dose calculation for photon-emitting brachytherapy sources with average energy higher than 50 keV: Report of the AAPM and ESTRO. Med Phys. 2012;39(5):2904-29. https://doi.org/10.1118/1.3703892 DOI: https://doi.org/10.1118/1.3703892

10. Beaulieu L, Carlsson Tedgren Å, Carrier J, Davis SD, Mourtada F, Rivard MJ, et al. Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation. Med Phys. 2012;39(10):6208-36. https://doi.org/10.1118/1.4747264 DOI: https://doi.org/10.1118/1.4747264

11. Guan F, Chen H, Draeger E, Li Y, Aydin R, Tien CJ, et al. Characterization of Gafchromic™ EBT4 film with clinical kV/MV photons and MeV electrons. Precis Radiat Oncol. 2023;7(2):84-91. https://doi.org/10.1002/pro6.1204 DOI: https://doi.org/10.1002/pro6.1204

12. Ainsbury EA, Dalke C, Hamada N, Benadjaoud MA, Chumak V, Ginjaume M, et al. Radiation-induced lens opacities: Epidemiological, clinical and experimental evidence, methodological issues, research gaps and strategy. Environ Int. 2021;146:106213. https://doi.org/10.1016/j.envint.2020.106213 DOI: https://doi.org/10.1016/j.envint.2020.106213

13. Thome C, Chambers DB, Hooker AM, Thompson JW, Boreham DR. Deterministic Effects to the Lens of the Eye Following Ionizing Radiation Exposure: is There Evidence to Support a Reduction in Threshold Dose? Health Phys. 2018;114(3):328-43. https://doi.org/10.1097/HP.0000000000000810 DOI: https://doi.org/10.1097/HP.0000000000000810

14. Christ G. White polystyrene as a substitute for water in high energy photon dosimetry. Med Phys.1995;22(12):2097-100. https://doi.org/10.1118/1.597652 DOI: https://doi.org/10.1118/1.597652

15. Howard ME, Herman MG, Grams MP. Methodology for radiochromic film analysis using FilmQA Pro and ImageJ. PLoS ONE. 2020;15(5):e0233562. https://doi.org/10.1371/journal.pone.0233562 DOI: https://doi.org/10.1371/journal.pone.0233562

16. Huang L, Gaballa H, Chang J. Evaluating dosimetric accuracy of the 6 MV calibration on EBT3 film in the use of Ir-192 high dose rate brachytherapy. J Appl Clin Med Phys. 2022;23(5):e13571. https://doi.org/10.1002/acm2.13571 DOI: https://doi.org/10.1002/acm2.13571

17. Bassi S, Cummins D, McCavana P. Energy and dose dependence of GafChromic EBT3-V3 film across a wide energy range. Phys Med. 2019;67:197-8. https://doi.org/10.1016/j.ejmp.2019.09.187 DOI: https://doi.org/10.1016/j.ejmp.2019.09.187

18. Berger MJ, Coursey JS, Zucker MA, Chang J. Stopping-power and range tables for electrons, protons, and helium ions (ESTAR, PSTAR, ASTAR); 2017. Online database. National Institute of Standards and Technology (NIST), Standard Reference Database 124. https://doi.org/10.18434/T4NC7P

Downloads

Published

2026-11-18

Issue

Section

Scientific articles

How to Cite

Study of ocular shielding in high dose rate brachytherapy for nasal skin lesions using Monte Carlo simulation and experimental measurements. (2026). Revista De Física Médica. https://doi.org/10.37004/sefm/2026.27.2.A003

Similar Articles

1-10 of 93