End-to-end testing of cranial radiosurgery on a Versa HD linear accelerator equipped with ExacTrac Dynamic
DOI:
https://doi.org/10.37004/sefm/2026.27.2.A001Keywords:
E2E test, radiosurgery, radiochromic films, anthropomor- phic phantom, ExacTracAbstract
This paper describes the end-to-end testing procedure implemented in our department for cranial radiosurgery delivered on an Elekta Versa HD linac equipped with an ExacTrac Dynamic image-guidance system, as well as the initial results obtained. The tests were carried out using an anthropomorphic phantom and radiochromic film dosimetry. Three single-isocenter modulated arc therapy plans were designed in the Monaco treatment planning system, with target volumes of different shapes and sizes. Treatments were delivered after positioning the phantom using ExacTrac X-ray images. The measured and planned doses were compared using gamma analysis with multiple criteria, and by evaluating the absolute dose difference at the center of the PTVs. Additionally, the geometric deviation of the dose delivery was estimated by comparing fiducial marker-based registration against registration based on optimizing dosimetric agreement. Observed deviations were generally within 1 mm and 5%, although areas with dose differences exceeding 5% of the local dose were found outside the PTVs in some cases.
References
1. García Romero A, Baeza Trujillo M, Teijeiro García A, Clemente Gutiérrez F, Morera Cano D, Hernández Masgrau V. Guía para el control de calidad y seguridad de los sistemas de planificación y planes de tratamiento de radioterapia externa. Rev Fis Med. 2024;25(1):123-82. https://doi.org/10.37004/sefm/2024.25.1.008 DOI: https://doi.org/10.37004/sefm/2024.25.1.008
2. Cirino E, Benedict SH, Dupre PJ, Halvorsen PH, Kim GGY, Reyhan ML, et al. AAPM-RSS Medical Physics Practice Guideline 9.b: SRS-SBRT. J Appl Clin Med Phys. 2025;26(4):e14624. https://doi.org/10.1002/acm2.14624 DOI: https://doi.org/10.1002/acm2.14624
3. Heukelom S, Hoffmans-Holtzer N, Marijnissen H, Nulens A, Pittomvils G, Raaijmakers E, et al. Process Management and Quality Assurance for Intracranial Stereotactic Treatment. Delft: The Netherlands Commission on Radiation Dosimetry; 2015. NCS Report 25. https://doi.org/10.25030/ncs-025 DOI: https://doi.org/10.25030/ncs-025
4. Mehrens H, Nguyen T, Edward S, Hartzell S, Glenn M, Branco D, et al. The current status and shortcomings of stereotactic radiosurgery. Neurooncol Adv. 2022;4(1):vdac058. https://doi.org/10.1093/noajnl/vdac058 DOI: https://doi.org/10.1093/noajnl/vdac058
5. Kry SF, Molineu A, Kerns JR, Alvarez P, Stingo F, Followill DS, et al. Institutional Patient-specific IMRT QA Does Not Predict Unacceptable Plan Delivery. Int J Radiat Oncol Biol Phys. 2014;90(5):1195-1201. https://doi.org/10.1016/j.ijrobp.2014.08.334 DOI: https://doi.org/10.1016/j.ijrobp.2014.08.334
6. Miura H, Ozawa S, Hosono F, Sumida N, Okazue T, Yamada K, et al. Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy. J Appl Clin Med Phys. 2016;17(6):312-322. https://doi.org/10.1120/jacmp.v17i6.6281 DOI: https://doi.org/10.1120/jacmp.v17i6.6281
7. Organismo Internacional de Energía Atómica. Determinación de la dosis absorbida en radioterapia externa: un código de práctica internacional para la dosimetría basada en patrones de dosis absorbida en agua. Viena: OIEA; 2025. Colección de Informes Técnicos Nº 398 (Rev. 1). Disponible en: https://www-pub.iaea.org/MTCD/publications/PDF/p15720-STIDOC010_398Rev1_web.pdf
8. Christ G. White Polystyrene as a Substitute for Water in High Energy Photon Dosimetry. Med Phys. 1995;22(12):2097-2100. https://doi.org/10.1118/1.597652 DOI: https://doi.org/10.1118/1.597652
9. Méndez I, Polšak A, Hudej R, Casar B. The Multigaussian method: A new approach to mitigating spatial heterogeneities with multichannel radiochromic film dosimetry. Phys Med Biol. 2018;63(17). https://doi.org/10.1088/1361-6560/aad9c1 DOI: https://doi.org/10.1088/1361-6560/aad9c1
10. Méndez I, Rovira-Escutia JJ, Casar B. A protocol for accurate radiochromic film dosimetry using Radiochromic.com. Radiol Oncol. 2021;55(3):369-378. https://doi.org/10.2478/raon-2021-0034 DOI: https://doi.org/10.2478/raon-2021-0034
11. Shariff M, Grigo J, Masitho S, Brandt T, Weiss A, Lambrecht U, et al. End-to-end testing for stereotactic radiotherapy including the development of a Multi-Modality phantom. Z Med Phys. 2024;34(3):477-484. https://doi.org/10.1016/j.zemedi.2022.11.006 DOI: https://doi.org/10.1016/j.zemedi.2022.11.006
12. Brodbek L, Kretschmer J, Büsing K, Looe HK, Poppe B, Poppinga D. Systematic end-to-end testing of multiple target treatments using the modular RUBY phantom. Biomed Phys Eng Express. 2022;8(1). https://doi.org/10.1088/2057-1976/ac3e37 DOI: https://doi.org/10.1088/2057-1976/ac3e37
13. Rousseau A, Stien C, Gouriou J, Bordy JM, Boissonnat G, Chabert I, et al. End-to-end quality assurance for stereotactic radiotherapy with Fricke-Xylenol orange-Gelatin gel dosimeter and dual-wavelength cone-beam optical CT readout. Phys Med. 2023;113. https://doi.org/10.1016/j.ejmp.2023.102656 DOI: https://doi.org/10.1016/j.ejmp.2023.102656
14. Magugliani G, Marranconi M, Liosi GM, Locatelli F, Gambirasio A, Trombetta L, et al. Pilot scale validation campaign of gel dosimetry for pre-treatment quality assurance in stereotactic radiotherapy. Phys Med. 2023;114. https://doi.org/10.1016/j.ejmp.2023.103158 DOI: https://doi.org/10.1016/j.ejmp.2023.103158
15. Bry V, Saenz D, Pappas E, Kalaitzakis G, Papanikolaou N, Rasmussen K. End to end comparison of surface-guided imaging versus stereoscopic X-rays for the SRS treatment of multiple metastases with a single isocenter using 3D anthropomorphic gel phantoms. J Appl Clin Med Phys. 2022;23(5). https://doi.org/10.1002/acm2.13576 DOI: https://doi.org/10.1002/acm2.13576
16. Dimitriadis A, Palmer AL, Thomas RAS, Nisbet A, Clark CH. Adaptation and validation of a commercial head phantom for cranial radiosurgery dosimetry end-to-end audit. Br J Radiol. 2017;90(1074):20170053. https://doi.org/10.1259/bjr.20170053 DOI: https://doi.org/10.1259/bjr.20170053
17. Calvo-Ortega JF, Pozo M, Moragues S, Casals J. Targeting accuracy of single-isocenter intensity-modulated radiosurgery for multiple lesions. Med Dosim. 2017;42(2):104-110. https://doi.org/10.1016/j.meddos.2017.01.006 DOI: https://doi.org/10.1016/j.meddos.2017.01.006
18. Xia Y, Adamson J, Zlateva Y, Giles W. Application of TG-218 action limits to SRS and SBRT pre-treatment patient specific QA. J Radiosurg SBRT. 2020;7(2):135-147. https://pubmed.ncbi.nlm.nih.gov/33282467/
19. Rose MS, Tirpak L, Van Casteren K, Zack J, Simon T, Schoenfeld A, et al. Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pre-treatment quality assurance. Med Phys. 2020;47(7):3153-3164. https://doi.org/10.1002/mp.14153 DOI: https://doi.org/10.1002/mp.14153
20. Knill C, Sandhu R, Halford R, Snyder M, Seymour Z. Commissioning cranial single-isocenter multi-target radiosurgery for the Versa HD. J Appl Clin Med Phys. 2021;22(4):108-114. https://doi.org/10.1002/acm2.13223 DOI: https://doi.org/10.1002/acm2.13223
21. Culcasi R, Baran G, Dominello M, Burmeister J. Stereotactic radiosurgery commissioning and QA test cases— A TG-119 approach for Stereotactic radiosurgery. Med Phys. 2021;48(12):7568-79. https://doi.org/10.1002/mp.15087 DOI: https://doi.org/10.1002/mp.15087
22. Brown TAD, Fagerstrom JM, Beck C, Holloway C, Burton K, Kaurin DGL, et al. Determination of commissioning criteria for multileaf-collimator, stereotactic radiosurgery treatments on Varian TrueBeam and Edge machines using a novel anthropomorphic phantom. J Appl Clin Med Phys. 2022;23(6). https://doi.org/10.1002/acm2.13581 DOI: https://doi.org/10.1002/acm2.13581
23. Lim SB, Kuo LC, Li T, Li X, Ballangrud AM, Lovelock M, et al. Comparative study of SRS end-to-end QA processes of a diode array device and an anthropomorphic phantom loaded with GafChromic XD film. J Appl Clin Med Phys. 2022;23(9). https://doi.org/10.1002/acm2.13747 DOI: https://doi.org/10.1002/acm2.13747