A traceability establishment for the 125-I IsoSeed I25.S06 source from Bebig at the hospital level

Authors

  • J Pérez-Calatayud
  • MJ Béjar
  • F Lliso
  • R Colmenares
  • J Fernández
  • M Andrassy
  • V Carmona
  • N Ferrer
  • J Vivanco

Abstract

The purpose of this work has been on one side to solve the traceability problem in hospitals for the I25.S06 seed users, already solved for other seeds, and on the other to design a holder in order to allow the measurement of the reference air-kerma rate for the strand by means of a well-type ionization chamber, widely used in brachytherapy in hospitals. At a joint calibration-measurement session of the involved institutions the following steps were performed: calibration of the HDR 1000 and SourceCheck detectors with a standard source from the Physikalisch-Technische Bundesanstalt (PTB); reference air-kerma rate measurements for the 5 seeds strand provided by Bebig‡ with the SourceCheck detector and with the HDR 1000 detector and the new insert holder by SIHO Spain); separation of the stranded seeds by the specialist from Bebig who dissolved the plastic; reference air-kerma rate measurement of each of the seeds of the strand with one of the of HDR 1000 detectors previously calibrated. We present the calibration factors obtained and the analysis of the uncertainties associated with the process.