Characterization of Siemens Biograph 6 PET by Monte Carlo simulation

Authors

  • Pedro Gallego Franco
  • Rocío García Marcos

Abstract

GAMOS simulation code based on Geant4 is a very powerful tool for the design and modeling optimization on Positron Emission Tomography (PET) systems. In order to obtain a proper image quality, it results to be extremely important determine the optimal activity which is going to be delivered. For this reason a study about the internal system parameters that affects image quality, such as scatter fraction (SF) and the count rate equivalent noise (NEC), has been carried out. The study involves the comparison of experimental measures on both parameters, with those obtained by Monte Carlo simulation of Siemens Pet Biograph 6 True Point with True V option. Based on simulation’s results, a paralizable dead-time model that adjusts, depending on the activity provided, the proper dead-time for scanner detectors. Also a study about the variation of this proper dead-time with the activity has been carried out.

Downloads

Issue

Section

Scientific articles

How to Cite

Characterization of Siemens Biograph 6 PET by Monte Carlo simulation. (2016). Revista De Física Médica, 16(3). https://revistadefisicamedica.es/index.php/rfm/article/view/181