A model for a CT dose profile in a standard PMMA head phantom

Authors

DOI:

https://doi.org/10.37004/sefm/2023.24.1.004

Keywords:

TC, model, dose profile

Abstract

Purpose: This work present a model for the dose profile deposited in a standard phantom of 16 cm of diameter and 15 cm of length (manufactured with "polymethyl methacrylate" - PMMA) in a CT or computed tomography.

Method: The model proposed is based on error and gaussian functions, where the most of parameters used has a simple physics justification. The data used for the development of this model have been obtained with a solid state probe.

Results and discussion: A good approach has been found to the experimental data with the proposed model, besides being of great versatility since a great variety of experimental measures have been used, with different equipment and acquisition parameters.

Conclusions: Our model adjust very well the deposition profiles of doses on a head phantom in its central axis. It can be considered as a basis for carrying out experimental corrections, dosimetric determinations and other considerations that require a theoretical dose profile model.

References

McCollough C, et al. “The measurement, reporting, and management of radiation dose in CT”. AAPM Task Group. 2008 Jan; 23(23):1-28.

Andersson Jonas, et al. “Estimating Patient Organ Dose with Computed Tomography: A Review of Present Methodology and Required DICOM Information: A Joint Report of AAPM Task Group 246 and the European Federation of Organizations for Medical Physics (EFOMP).” (2019).

F.R. Verdun, et al. “Image quality in CT: From physical measurement to model observers” Physica Medica, Volume 31, Issue 8, 2015, 823-843.

Shope TB, Gagne RM, Johnson GC. “A method for describing the doses delivered by transmission x-ray computed tomography”. Med Phys. 1981 Jul-Aug; 8(4):488-95.

Gagne R M. “Geometrical aspects of computed tomography: sensitivity profile and exposure profile”. Med Phys. 1989 Jan-Feb; 16(1):29-37.

Weir, Victor J. “A model of TC dose profiles in banach space; with applications to TC dosimetry.” Physics in Medicine & Biology 61.13 (2016): 5020.

International Electrotechnical Commission (IEC). Medical Electrical Equipment. Part 2-44: Particular requirements for the safety of x-ray equipment for computed tomography. IEC publication No. 60601-2- 4. Ed. 2.1: International Electrotechnical Commission (IEC) Central Office: Geneva, Switzerland, 2002.

Choirul Anam, et al. “Comparison of central, peripheral, and weighted size-specific dose in TC.” Journal of X-ray Science and Technology 28.4 (2020): 695-708.

Choirul Anam, et al. “Scatter index measurement using a CT dose profiler”. Journal of Medical Physics and Biophysics, Vol. 4, No.1, August 2017.

Zhang D, et al. “Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms”. Med Phys. 2009 Mar; 36(3):1025-38.