Reduction in the number of measurements needed in order to characterize a system of intraoperative radiotherapy applicators

Authors

  • MA de la Casa de Julián
  • E Cabello Murillo
  • F Clemente Gutiérrez
  • J García Ruiz-Zorrilla
  • R Díaz Fuentes
  • A Ferrando Sánchez
  • J Castro Novais

Abstract

Intraoperative radiotherapy allows the administration of a high radiation dose, in a single session, to an exposed tumor or a surgical site. Radiation is delivered by means of a set of applicators, where each applicator has a set of bevels with different angles. The dosimetric characterization of the set of applicators and bevels requires the performance of a large number of measurements. In this work, we present a procedure for the dosimetric characterization of a set of intraoperative radiotherapy applicators that significantly reduces the number of dosimetric measurements required, while preserving a satisfactory level of accuracy when compared to the complete series of measurements which in principle would be required.

Downloads

Issue

Section

Scientific articles

How to Cite

Reduction in the number of measurements needed in order to characterize a system of intraoperative radiotherapy applicators. (2011). Revista De Física Médica, 12(2). https://revistadefisicamedica.es/index.php/rfm/article/view/54