Dosimetric analysis and verification measurement of VMAT treatment in twins linacs

Authors

  • Alba Obesso de Diego Servicio de Radiofísica y Radioprotección, Hospital Universitario la Paz.
  • Carlos Ferrer Gracia Servicio de Radiofísica y Radioprotección, Hospital Universitario la Paz.
  • Concepción Huertas Martínez Servicio de Radiofísica y Radioprotección, Hospital Universitario la Paz.
  • Ana Morcillo García Servicio de Radiofísica y Radioprotección, Hospital Universitario la Paz.
  • Rodrigo Plaza Núñez Servicio de Radiofísica y Radioprotección, Hospital Universitario la Paz.

DOI:

https://doi.org/10.37004/sefm/2021.22.1.002

Keywords:

Beam matching, dosimetric verification, VMAT treatments, gamma analysis

Abstract

The management of radiotherapy treatments is improved if there is the possibility of treating a patient indistinctly in different linacs without having to make changes to treatment plan. To make this possible, the dosimetric and mechanical characteristics of the linacs must be the same. The purpose of this work was to verify that the Elekta Synergy and Infinity linacs, both with 6 MV photon beam, flattening filter and Agility treatment head, can be considered twins. For this purpose, different dosimetric parameters were measured in the Infinity which were compared later with the Synergy linac acceptance tests. Finally, the quality of the VMAT (Volumetric Modulated Arc Therapy) treatments was evaluated through gamma analysis (3 %/ 2 mm). A good concordance has been observed regarding the dosimetric parameters analyzed and the gamma analysis of the VMAT treatments. Specifically, the difference obtained was < 2 % in all the beam quality parameters and < 3 % in all the VMAT treatments. The possibility of using the same modeling for both linacs is confirmed, which allows to treat in one or the other interchangeably.

References

INTERNATIONAL ATOMIC ENERGY AGENCY, Radiation Protection in the Design of Radiotherapy Facilities, Safety Reports Series No. 47, IAEA, Vienna (2006).

Sjöström D, Bjelkengren U, Ottosson W, Behrens CF. A beam-matching concept for medical linear accelerators. Acta Oncol 2009;48(2):192-200.

Attalla EM, Abou-Elenein HS, Ammar H, Eldesoky I. Dosimetric evaluation of a beam matching procedure. Chinese-German J Clin Oncol 2014;13(2):89-93.

Bhangle JR, Narayanan VK, Kumar NK, Vaitheeswaran R. Dosimetric analysis of beam-matching procedure of two similar linear accelerators. J Med Phys 2011;36:176–80.

Fenoglietto P, Khodri M, Nguyen D, Josserand-Pietri F, Aillères N. Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study. Radiat Oncol 2016 Jan 14;11:2.

Sangaiah Ashokkumar KM, Ramalingam K, Karthikeyan K, Jagadheeskumar N. Dosimetric Validation of Volumetric Modulated Arc Therapy with Three 6MV Beam-Matched Linear Accelerators. Asian Pac J Cancer Prev.: APJCP 2017;18(12): 3439-3444.

Treutwein M, Härtl PM, Gröger C, Katsilieri Z, Dobler B. Linac twins in radiotherapy. Evolution of Ionizing Radiation Research 2015; Rijeka: InTech:171-86.

Xu Z, Warrell G, Lee S, Colussi V, Zheng Y, Ellis R, et al. Assessment of beam-matched linacs quality/accuracy for interchanging SBRT or SRT patient using VMAT without replanning. J Appl Clin Med Phys 2019;20(1):68-75.

Latala A, Fujak E, Walewska A, Kukołowicz P. The comparison of VMAT test results for Clinac 2300C/D and TrueBeam accelerators. Med Dosim 2020 Autumn;45(3):219-224.

Palmans H, Andreo P, Huq MS, Seuntjens J, Christaki KE, Meghzifene A. Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA-AAPM international Code of Practice for reference and relative dose determination. Med Phys 2018;45(11):e1123-e1145.

Palta JR. Linear Accelerator Acceptance Testing and Commissioning. Med Phys 2003;30(6):1356-1357.

Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, et al. Task Group 142 report: quality assurance of medical accelerators. Med Phys 2009;36(9):4197-4212.

Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218. Med Phys 2018;45(4):e53-e83.

Tessier F, Kawrakow I. Effective point of measurement of thimble ion chambers in megavoltage photon beams. Med Phys. 2010 Jan;37(1):96-107.

Andreo P, Burns DT, Hohlfeld K, Huq MS, Kanai T, Laitano F, et al. IAEA TRS-398: Absorbed dose determination in external beam radiotherapy. Vienna, Austria: International Atomic Energy Agency. 2000.

Gonzalez Lopez A, Vera Sanchez JA, Lago-Martin JD. Small fields measurements with radiochromic films. J Med Phys 2015 Apr-Jun;40(2):61-7.

Kinsella P, Shields L, McCavana P, McClean B, Langan B. Determination of MLC model parameters for Monaco using commercial diode arrays. J Appl Clin Med Phys 2016 Jul 8;17(4):37-47.

Casar B, Gershkevitsh E, Mendez I, Jurković S, Huq MS. A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams. Med Phys 2019 Feb;46(2):944-963.

Cohen J. Statistical power analysis for the behavioral sciences. 2 ed. New York: Psychology Press; 1988.

García-Garduño OA, Lárraga-Gutiérrez JM, Rodríguez-Villafuerte M, Martínez-Dávalos A, Celis MA. Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom. Radiother Oncol 2010 Aug;96(2):250-3.

Morales JE, Butson M, Crowe SB, Hill R, Trapp JV. An experimental extrapolation technique using the Gafchromic EBT3 film for relative output factor measurements in small x-ray fields. Med Phys 2016 Aug;43(8):4687.

Wilcox EE, Daskalov GM. Evaluation of GAFCHROMIC EBT film for Cyberknife dosimetry. Med Phys 2007 Jun;34(6):1967-74.

Li Y, Chen L, Zhu J, Wang B, Liu X. A quantitative method to the analysis of MLC leaf position and speed based on EPID and EBT3 film for dynamic IMRT treatment with different types of MLC. J Appl Clin Med Phys 2017;18(4):106-115.

McKenzie EM, Balter PA, Stingo FC, Jones J, Followill DS, Kry SF. Reproducibility in patient-specific IMRT QA. J Appl Clin Med Phys 2014;15(3):4741.

Downloads

Published

2021-04-02

Issue

Section

Scientific articles

How to Cite

Dosimetric analysis and verification measurement of VMAT treatment in twins linacs. (2021). Revista De Física Médica, 22(1), 27-37. https://doi.org/10.37004/sefm/2021.22.1.002

Similar Articles

31-40 of 107

You may also start an advanced similarity search for this article.